The measured value is then no longer displayed, scanned, and output but the
programming is retained intact.
Reactivating the measuring channel:
Function: RANGE ´R´ keys : ,
PROG
PROG
If the channel was previously deactivated, then it is now reactivated with all pro-
gramming values intact. However, if the channel is already active, then with
this key combination all its programming values are deleted (i.e. as per se-
lecting the measuring range).
7.7.1 Special meas. ranges, linearization,
multi-point calibration
Thanks to the new ALMEMO special connectors with extra memory for additio-
nal data (larger EEPROM, code E4) the following tasks can now for the first ti-
me be elegantly performed :
1. Provision of special measuring ranges with internal characteristic (s. 10.3.9)
2. Linearization of signals for voltage, current, resistance, or frequency set by
the user
3. Multi-point calibration of all sensors
The ALMEMO 2890-9 can as standard evaluate all appropriately programmed
connectors. With the special KL variant it is also possible to convert measuring
signals into appropriate display values in line with a characteristic of up t 35
support values. These support points are programmed to the EEPROM in the
ALMEMO connector using the AMR-Control software. During a measuring ope-
ration the measured values are interpolated between these on a linear basis.
When correcting non-linear sensors (e.g. with Pt100 or thermocouple sensors)
initially the original characteristics are considered; then only the deviations are
interpolated on a linear basis and inserted.
If a channel with a characteristic is deactivated or programmed with a different
range, the characteristic is subsequently reactivated by programming the spe-
cial range ´Lin ´ by means of the keyboard or command “B99”.